Serving the fNIRS community
The properties that make fNIRS attractive in neuroscience are even more attractive at the bedside. A clinical monitor should be safe, noninvasive, continuous, and small enough to use in an incubator, an operating room, or an intensive care unit, where moving a patient to a scanner is risky or impossible. Near-infrared spectroscopy meets all of these requirements, and it was used on sick newborns within a few years of its first demonstration in the brain. Since then, it has been applied to monitor oxygenation in neonatal and adult intensive care and during surgery, and to study and support patients with stroke, epilepsy, and psychiatric disorders.
Clinical use, however, sets a higher bar than research. A measurement that works, and even one that changes in the expected direction in a group of patients, is not yet a measurement that helps an individual patient. In this lesson, we look at the main clinical areas with that distinction in mind, and we try to place each one at its real level of maturity. We also discuss what blood flow measurements, such as those of diffuse correlation spectroscopy (DCS), add to oxygenation at the bedside, because many clinical questions are, at their core, questions about flow.
Clinical NIRS is used in two different ways, and they are judged by different standards. The first is monitoring: a continuous measurement of oxygenation, usually StO₂ from a clinical oximeter (Lesson 4.3), that warns the clinical team when the brain or another organ may be receiving too little oxygen. The second is functional assessment: the same task-based measurements we have discussed throughout the course, used to characterize the brain of a patient, for example to locate a function before surgery, to follow recovery, or to support a diagnosis. For both, the path from a measurement to a benefit for patients has several steps, and each must hold.
💡 The Chain from Measurement to Outcome
A clinical measurement improves outcomes only if every link holds:
Most clinical NIRS research has addressed the first two links. The neonatal trials of Section 2 are among the few that tested the last one.
The newborn was the first clinical target of cerebral NIRS, for good reasons: the thin scalp and skull let light reach the brain easily (Lesson 3.5), the brain of a sick preterm infant is vulnerable to both too little and too much oxygen, and the alternatives are limited, since moving an unstable infant to a scanner is risky. Brazy et al. (1985), with Jöbsis among the authors, monitored sick preterm infants at the bedside and observed changes in hemoglobin oxygenation, blood volume, and the oxidation state of cytochrome oxidase during episodes of hypoxia and bradycardia. Edwards et al. (1988) went a step further and used NIRS to quantify cerebral blood flow at the cot side in ill newborns, showing how much it could vary between infants and with treatment.
Clinical cerebral oximeters, which report StO₂ continuously, are now used in many neonatal units and during cardiac surgery in children and adults. Whether their use improves outcomes was tested in a series of trials in extremely preterm infants, born before 28 weeks of gestation. In the phase II SafeBoosC trial, Hyttel-Sørensen et al. (2015) randomized 166 infants to cerebral oximetry with a dedicated treatment guideline, or to blinded monitoring with standard care, during the first 72 hours of life. Oximetry worked as intended on the physiological level: the burden of cerebral hypoxia and hyperoxia was reduced by 58%. The trial was not designed to detect differences in clinical outcomes, and none were found. The phase III trial, SafeBoosC III, was designed to answer that question. Hansen et al. (2023) randomized 1601 infants at 70 sites in 17 countries and found that death or severe brain injury occurred in 35.2% of infants managed with cerebral oximetry and in 34.0% of those receiving usual care, with no significant difference.
This is one of the most important results in clinical NIRS, and it deserves a careful reading. It does not show that cerebral oximetry is inaccurate or useless at the bedside, since the phase II trial showed that, combined with a treatment guideline, it stabilized cerebral oxygenation. It shows that, in this population, during these 72 hours, and with the treatments that were available to correct low or high values, stabilizing cerebral oxygenation did not reduce death or severe brain injury. Any of the links in the chain above might be responsible: the outcome may be driven mostly by factors that oxygenation does not capture, or the interventions available may not be the right ones. Other populations, time windows, or treatment strategies may give different answers, and that is precisely why they have to be tested rather than assumed.
🎍 MUST-READ PAPER
Hansen, M.L., Pellicer, A., Hyttel-Sørensen, S., et al. (2023). “Cerebral oximetry monitoring in extremely preterm infants.” New England Journal of Medicine, 388(16), 1501–1511.
The SafeBoosC III trial, the largest randomized test of NIRS-guided care to date. Read it together with the phase II trial (Hyttel-Sørensen et al., 2015) to see the difference between improving a physiological measure and improving a clinical outcome, and how a well-designed trial distinguishes the two.
After a severe brain injury, such as trauma, subarachnoid hemorrhage, or a large stroke, the brain is at risk of secondary injury from inadequate blood flow, and the intensive care team works to keep its perfusion within safe limits. A central concept here is cerebral autoregulation: the ability of the healthy brain to keep its blood flow nearly constant when blood pressure changes. When autoregulation is impaired, blood flow follows blood pressure passively, and both low and high pressures become dangerous. Assessing autoregulation continuously, at the bedside, would allow blood pressure targets to be set for each patient. NIRS offers one way to do this. Brady et al. (2007), in an infant animal model, computed a moving correlation between cerebral perfusion pressure and the StO₂ of a clinical oximeter: when perfusion pressure fell below the limit of autoregulation, oxygenation began to follow pressure, and the correlation rose. Indices of this kind are now studied in adult and pediatric intensive care.
Notice, however, that an oxygenation-based index uses StO₂ as a stand-in for blood flow. Oxygenation depends on flow, but also on oxygen consumption, arterial saturation, and hemoglobin concentration (Lesson 1.3), any of which can change in a critically ill patient. This is where the blood flow measurements of Lesson 4.4 become valuable. Durduran et al. (2009) combined DCS and NIRS to monitor patients with acute ischemic stroke while the head of the bed was lowered from 30° to −5°, and found changes in blood flow consistent with impaired autoregulation in the hemisphere with the infarct. Kim et al. (2010) measured patients in a neurointensive care unit during changes in blood pressure and CO₂ and compared DCS with xenon-enhanced CT, a clinical method for measuring cerebral blood flow: relative changes in flow measured by the two methods were well correlated (r = 0.73), while the NIRS hemoglobin signals agreed with flow only moderately.
With flow and oxygenation measured together, the hybrid instrument can also estimate oxygen metabolism, as we saw in Lesson 4.4. This matters clinically because the same fall in oxygenation can mean very different things. It can mean that flow has fallen and the brain is at risk of ischemia, or that consumption has risen, as during a seizure, or that the arterial blood carries less oxygen. Each calls for a different response. Durduran et al. (2010) used a hybrid DCS and NIRS system in 33 newborns with congenital heart disease, validated the DCS flow changes against arterial spin labeling MRI (R = 0.7), and estimated changes in cerebral oxygen metabolism, in a population in which the brain is at risk before, during, and after cardiac surgery. Buckley et al. (2009) showed that DCS could be used in very preterm infants in the intensive care unit, with blood flow that correlated with Doppler ultrasound measurements of blood flow velocity.
💭 Pause and Think
In a ventilated patient in the intensive care unit, the cerebral oximeter reading falls from 70% to 58% over 20 minutes. Arterial oxygen saturation, measured with a pulse oximeter, has not changed. List the possible explanations, and decide which additional measurement would help you most to tell them apart.
Think about blood pressure and autoregulation, arterial CO₂ (Lesson 3.5), oxygen consumption during a seizure, hemoglobin concentration, changes in the scalp under the probe (Lesson 4.3), and what a simultaneous measurement of blood flow would add.
DCS remains, for now, mostly a research tool in clinical settings. The limits discussed in Lesson 4.4 (few photons, few channels, short separations, cost, and blood flow indices that are easier to compare within a patient than between patients) weigh even more in an intensive care unit, where measurements must run for days and the scalp can change with temperature, swelling, and the pressure of the probe. The newer speckle-based techniques of Lesson 4.4 aim at exactly these limits. Clinical NIRS oximetry, for its part, would benefit from the absolute and depth-resolved measurements of Lesson 4.3, since a single number from the forehead combines brain and extracerebral tissue in proportions that differ between patients.
Recovery after stroke takes place over weeks to months, much of it in rehabilitation gyms rather than in imaging centers, and it involves exactly the activities that a scanner cannot accommodate, such as standing, walking, and reaching. This makes rehabilitation a natural setting for fNIRS. Miyai et al. (2002) extended their study of gait (Lesson 6.1) to six patients who could not walk after severe stroke, recording them while they walked on a treadmill with body-weight support and the help of physical therapists. Walking increased activation of the premotor cortex in the affected hemisphere, and activation and gait performance were both greater with a therapy technique that facilitated the movement of the paralyzed leg, suggesting that the premotor cortex contributes to the recovery of gait.
fNIRS can also be part of the therapy. In neurofeedback, the patient sees a signal derived from their own brain activity in real time and learns to increase it. Mihara et al. (2013) randomized 20 patients with stroke to motor imagery training with real feedback of their cortical NIRS signals or with sham feedback, in addition to standard rehabilitation, and found greater improvement in hand and finger function in the group that received real feedback. This was a small pilot study, and larger trials are needed before such approaches become standard care, but it illustrates a use of fNIRS that goes beyond measurement. We return to real-time uses of fNIRS, including brain-computer interfaces, in Lesson 6.4.
Stroke also brings specific caveats. A lesion changes the optical properties and the blood supply of the tissue under the probe, and the neurovascular coupling of the surviving tissue may itself be impaired (Lesson 3.1), so a smaller response in the affected hemisphere does not necessarily mean less neural activity. Patients may also move differently, tire sooner, and have different blood pressure responses than controls, which makes the physiological confounds of Lesson 3.5 harder to match between groups.
Two questions in epilepsy have attracted fNIRS. The first arises before epilepsy surgery, when the surgical team needs to know which hemisphere is dominant for language. Traditionally, this was assessed with the Wada test, in which an anesthetic is injected into one carotid artery to silence one hemisphere at a time. Watanabe et al. (1998) showed that a 24-channel fNIRS system could identify the side of language dominance during a word-generation task, in agreement with the Wada test in six patients with epilepsy. fMRI has since become the main noninvasive alternative, but fNIRS remains useful in children and in patients who cannot be scanned.
The second question concerns the seizures themselves. Seizures are unpredictable, often involve movement, and can last minutes, so studying their hemodynamics requires long, continuous recordings in a hospital bed, a setting where fNIRS combines naturally with EEG, the clinical standard for detecting seizures. Nguyen et al. (2012) recorded continuous EEG and fNIRS in patients with temporal lobe epilepsy and captured eight seizures. Over the temporal lobe where the seizures began, oxygenation increased and was then followed by a deoxygenated state, and changes also appeared over the opposite temporal lobe and in remote frontal and parietal regions, even while the EEG still showed epileptic activity limited to the temporal lobe. These recordings show how fNIRS can add a vascular and metabolic view to the electrical view of EEG. They also show its limits: seizures often begin in deep structures, such as the hippocampus, that fNIRS cannot reach (Lesson 1.4), so what it records is the cortical consequence of the seizure, mixed with the large systemic changes in heart rate, blood pressure, and breathing that seizures cause.
Psychiatric diagnosis relies on interviews and observation, without the objective biological tests that guide much of the rest of medicine, and a measurement that could support it would be valuable. Because fNIRS is quick, quiet, and can be performed in a clinic, it has been explored for this purpose, particularly in Japan. Suto et al. (2004) measured the frontal and temporal cortices of patients with depression, patients with schizophrenia, and healthy controls during a verbal fluency task, and found that the time course of the HbO₂ increase differed between the groups even when task performance was similar. Takizawa et al. (2014) then built a diagnostic support system based on this approach and tested it in 673 patients with depressive symptoms and 1007 healthy volunteers at seven hospitals. Using a threshold defined at one site and validated at the other six, the frontal hemodynamic pattern classified patients with major depressive disorder correctly in 74.6% of cases, and patients with bipolar disorder or schizophrenia in 85.5% of cases.
This work is a good example of how a clinical application should be built: a large sample, several sites, and a threshold validated on data that were not used to choose it. It is also a reminder of how much remains open. An accuracy of 75% means that one patient in four would be misclassified, which limits the role of the test to supporting, not replacing, clinical judgment. And the measurement is taken from the forehead during a speaking task, the situation in which scalp blood flow, breathing, and autonomic arousal contribute most to the signal (Lessons 3.5 and 6.1), all of which can differ between psychiatric groups and with medication. The authors themselves noted that the effects of systemic variables, such as indices of the autonomic nervous system, needed to be clarified.
⚠ A Common Mistake to Avoid
It is tempting to describe a significant difference between a patient group and a control group as a “biomarker” of the disease. A group difference only shows that the averages differ; the distributions of individual patients and controls can still overlap almost completely, and a measure that separates group means very reliably can still classify individuals barely better than chance.
A clinical test must be evaluated at the level of the individual (sensitivity, specificity, and how they depend on the threshold), in data that were not used to build it, in the population in which it will be used, and against the confounds that differ between groups, such as medication, arousal, systemic physiology, and the thickness of the scalp. Only then can it be tested for what really matters: whether using it improves care.
🔬 Explore It Yourself
Change the difference between patients and controls and the number of participants. Watch the p-value of the group comparison shrink as the sample grows, while the proportion of individuals that could be classified correctly does not move at all.
Table 1 summarizes the clinical applications of this lesson. The pattern is consistent: NIRS is easy to bring to the patient, and in every area it has produced physiological insight that would be hard to obtain otherwise. Its main clinical use today is oxygenation monitoring, in neonatal and cardiac care, where it is widely used even though the largest trial so far has not shown that it improves outcomes. Functional applications have reached the stage of validation in patient groups, and a few have been tested in randomized studies, but none is yet a routine part of care for individual patients. The gap is not in the technology alone. It lies in the steps that connect a measurement to a decision, and those steps require clinical trials, not just better instruments.
| Area | Why NIRS here | Where it stands |
|---|---|---|
| Neonatal and cardiac care | Continuous cerebral oxygenation at the bedside in patients who cannot be moved | Widely used for monitoring; in extremely preterm infants, NIRS-guided care stabilized oxygenation but did not improve outcomes in a large trial |
| Neurocritical care | Continuous assessment of autoregulation, flow, and metabolism without transport to a scanner | Oxygenation-based autoregulation indices under clinical study; DCS validated against clinical flow methods, mostly in research settings |
| Stroke and rehabilitation | Measurements during real movement and therapy, repeated over recovery | Insight into recovery mechanisms; neurofeedback promising in small randomized studies |
| Epilepsy | Language lateralization without injection or scanner; long recordings with EEG | Lateralization agrees with the Wada test in small series; seizure hemodynamics a research tool, limited by depth |
| Psychiatry | A quick, quiet measurement that can be performed in a clinic | Multisite validation of diagnostic support for depressive symptoms; accuracy and systemic confounds still limit its use for individuals |
Table 1. Clinical applications of NIRS discussed in this lesson, why NIRS is used in each, and their level of maturity.
📌 Key Takeaways
📚 Further Reading & Key References
Society for Functional Near-Infrared Spectroscopy.
Content is for educational purposes.