Introduction to fNIRS Continuous-Wave fNIRS Systems
Introduction to fNIRS
0% Completed

Inside a Continuous-Wave fNIRS System

In Units 2 and 3, we treated the fNIRS instrument as a device that delivers light to the head and reports how much of it comes back, and we focused on what that measurement means physically and physiologically. In this unit, we look at the instrument itself. Understanding what happens inside the hardware is useful even if you never build one: it helps you choose a system for your question, set it up in a way that gives good data, recognize when the hardware (rather than the physiology) is limiting your measurement, and read the methods sections of other studies with a critical eye.

We will start with a map of the four families of instruments you will find in the literature, so that you know where the rest of the unit is heading. We will then look in detail at the most common family, the continuous-wave (CW) system: how much light actually comes back from the head and what that implies for noise, the light sources and detectors used to generate and measure it, how the instrument tells its channels apart, and what changes when the system moves from a laboratory bench to a wearable device. We close with the safety aspects that every user should know. Lesson 4.2 then turns to designing and running an experiment with this kind of system.

1. A Map of fNIRS Instruments

All fNIRS instruments do the same basic thing: they send near-infrared light into the head through the scalp and detect the fraction that returns to the surface a few centimeters away. Where they differ is in which property of the returning light they measure. A CW system measures only its intensity; frequency-domain (FD) and time-domain (TD) systems also measure how long the light took to travel through the tissue; and speckle-based systems measure how the light fluctuates over time because of moving red blood cells. As we saw in Lessons 1.3 and 2.5, this choice decides what the instrument can tell you about the tissue.

Family What it measures What that gives you Lesson
Continuous wave (CW) The intensity of the detected light Relative changes in HbO₂ and HbR. Simple, affordable, and by far the most common 4.1, 4.2
Frequency domain (FD) Intensity and phase of light modulated at radio frequencies Absolute absorption and scattering, and therefore absolute concentrations and StO₂ 4.3
Time domain (TD) The distribution of arrival times of photons from very short light pulses Absolute values, plus depth selectivity from late-arriving photons 4.3
Speckle-based (DCS, SCOS) Temporal fluctuations of coherent light caused by moving red blood cells An index of blood flow, which absorption-based methods cannot provide 4.4

These families are not competitors; rather, they trade simplicity and cost for information. A CW system is the least informative of the four (it cannot separate absorption from scattering, and therefore can only report relative changes), but it is also the simplest to build, the easiest to miniaturize, and the one that most fNIRS studies use. Much of what we discuss in this lesson and in the next one, such as sources, detectors, optodes, safety, and experimental design, also applies to the other three families, which we will discuss in Lessons 4.3 and 4.4.

2. From Light to Hemoglobin: The Signal Chain

Conceptually, a CW system performs a simple task. It shines light of at least two wavelengths into the head, measures the intensity of the light that returns at each detector, and repeats this measurement several times per second for every source-detector pair of interest. Almost everything else in the instrument exists to perform this task accurately, for many channels at the same time, and without the channels interfering with each other or with the light in the room.

Block diagram of a CW fNIRS system: encoded light sources, optodes on the head, detectors, electronics, raw intensity data, and conversion to HbO2 and HbR changes

Figure 1. The signal chain of a CW fNIRS system. The instrument records only the intensity of the returning light over time; converting it to concentration changes requires the modified Beer-Lambert law and an assumed DPF (Unit 2).

It is worth being clear about what the instrument actually measures. The raw output of any CW system is a set of intensity time series, one per channel and per wavelength. Optical densities and hemoglobin concentration changes are computed from these intensities afterwards, using the modified Beer-Lambert law, tabulated extinction coefficients, and an assumed DPF (Unit 2), and many systems do this computation in real time for display. We strongly recommend keeping the raw intensities, and not only the hemoglobin changes derived from them: many signal quality checks work directly on the raw data (Lesson 4.2 and Unit 5), and a different DPF, a corrected source wavelength, or a new analysis approach can only be applied later if the raw data are still available.

3. How Much Light Comes Back?

Before discussing sources and detectors, it helps to know the numbers they have to deal with. We can estimate how much light returns to the surface with the same diffusion theory introduced in Unit 2: for a homogeneous semi-infinite medium, Farrell et al. (1992) derived an analytical expression for the diffuse reflectance as a function of the distance from the source. Using typical optical properties of the adult head (μa ≈ 0.15 cm⁻¹ and μs′ ≈ 10 cm⁻¹), a source injecting 10 mW of light at 800 nm, and a detector collecting light over about 7 mm² (roughly the area of a 3 mm fiber bundle), the model gives the following estimates:

Source-detector separation Fraction of injected light collected Detected power (10 mW source) Photons per second
1 cm ~10⁻³ ~8 µW ~3 × 10¹³
2 cm ~3 × 10⁻⁵ ~0.3 µW ~10¹²
3 cm ~10⁻⁶ ~10 nW ~5 × 10¹⁰
4 cm ~10⁻⁷ ~1 nW ~4 × 10⁹
5 cm ~7 × 10⁻⁹ ~70 pW ~3 × 10⁸

These are idealized numbers from a homogeneous model, and in a real measurement the detected light is usually lower, often by an order of magnitude or more, because of losses in the fibers and optical couplings, hair under the optodes, and the layered structure of the head. Two features, however, hold in any case. First, the detected light falls steeply with distance: in this model, by a little more than an order of magnitude for every additional centimeter. Second, the range of intensities an instrument must handle is enormous: a short channel at 1 cm and a long channel at 4 cm on the same probe differ by about four orders of magnitude, and both must be measured at the same time.

These numbers also tell us something about noise. Photons arrive at random, so even a perfect detector counting N photons has a statistical uncertainty of about √N, known as shot noise; its relative size, 1/√N, falls as more light is detected. At 3 cm, a sample of 0.1 s (i.e., a sampling rate of 10 Hz) collects billions of photons, and the relative shot noise is on the order of 10⁻⁵. For comparison, a typical functional response changes the detected intensity by roughly 1% (an optical density change on the order of 0.01–0.02). In other words, at standard separations the fundamental photon statistics are not what limits fNIRS. In practice, the limits come from the noise added by detectors and electronics when the light is weak (at long separations, or through dark or dense hair), from ambient light, from poor optode contact, and, most of all, from the systemic physiology and motion discussed in Lessons 2.3 and 3.5. The steep loss of light with distance is also why simply moving the detector further away, to reach deeper, quickly becomes impractical.

💭 Pause and Think

You want more sensitivity to the brain and consider moving your detectors from 3 cm to 4.5 cm. Using the table above, by roughly how much would the detected light fall, and what else would you need to change to keep the same signal quality?

Think about the source power (and its safety limits), the detector, the integration time, and what the extra depth actually buys you (Lesson 2.5).

🔬 Explore It Yourself

Check your answer to the question above, then explore further. Move the separation, change the tissue and the instrument, and use the “extra losses” slider to see how quickly a channel that looks comfortable on paper can approach the limits of detection.

4. Light Sources: LEDs and Laser Diodes

CW systems use one of two types of light source, light-emitting diodes (LEDs) or laser diodes, typically delivering a few milliwatts to a few tens of milliwatts of optical power per source into the tissue. The upper end of this range is set by safety rather than by technology (Section 8). The choice between the two types has consequences for both the hardware and the data.

Light-emitting diodes (LEDs) Laser diodes
Spectral width Broad, typically a few tens of nanometers Narrow, typically a few nanometers or less
Delivery Usually placed directly on the skin Usually coupled into optical fibers
Strengths Small, inexpensive, and power-efficient; well suited to wearable and fiberless designs Well-defined wavelength; high power can be coupled efficiently into thin fibers
Things to watch Broad spectrum makes the effective extinction coefficients less certain; heat conducted from the LED to the skin Eye safety; fibers add weight and restrict movement

The spectral width matters more than it may seem at first. The modified Beer-Lambert law uses the extinction coefficients of HbO₂ and HbR at the wavelength of the source (Lesson 2.3). With a laser diode, that wavelength is well defined. With an LED, the emitted light is a mixture of wavelengths spanning a few tens of nanometers, and hemoglobin absorption changes appreciably over that range, especially below 800 nm, where the HbR spectrum is steep. The appropriate extinction coefficient is then an average weighted by the LED’s emission spectrum, rather than the value at its nominal peak. In addition, the peak wavelength of both LEDs and laser diodes shifts with temperature (by a fraction of a nanometer per degree), and the emitted power changes as the device warms up. This is one reason why instruments should be switched on some time before a measurement, why slow drifts at the start of a recording should be interpreted with care, and why the actual source wavelengths (and, for LEDs, their spectra) should be reported in publications.

5. Detectors: Turning Light into a Signal

Given the numbers in Section 3, a detector for fNIRS has two competing jobs. It must be sensitive enough to measure the nanowatts (or less) that return from a long channel with low added noise, and it must not saturate when the microwatts from a short channel arrive on the same probe. Four types of detector are in common use, and they balance these requirements differently.

Detector Internal gain Efficiency in the NIR (typical) Strengths and limitations
Silicon photodiode None High (roughly 50–80%) Robust, inexpensive, very wide dynamic range; without gain, the amplifier noise limits performance at low light
Avalanche photodiode (APD) Moderate (tens to hundreds) High More sensitive than a photodiode at low light; gain depends on voltage and temperature, and the gain process adds some noise
Silicon photomultiplier (SiPM) Very high (~10⁵–10⁶) Moderate (often ~5–20%) Compact, low voltage, and sensitive to single photons; well suited to wearable high-density systems, but saturates at high light levels
Photomultiplier tube (PMT) Very high (~10⁶) Low to moderate (often a few percent to ~10%) Extremely sensitive; bulky, needs high voltage, and can be damaged by strong light; used mostly in fiber-based systems

The values in the table are indicative, since each class covers many devices. The important point is the trade-off between efficiency (the fraction of arriving photons that produce a signal) and gain (how much each detected photon is amplified inside the detector). A silicon photodiode converts most of the arriving near-infrared photons into electrons, but it provides no amplification, so at very low light levels the noise of the following amplifier dominates. Detectors with internal gain (APDs, SiPMs, and PMTs) lift the signal above that electronic noise, which is why they are preferred when the light is weak, but they are more easily saturated and, in the case of PMTs, more fragile. The detection area also matters: a larger area, or a fiber bundle with a wider aperture, collects more light, at the cost of spatial precision and, for fiberless designs, of size and weight on the head.

For the user, all of this translates into a small number of practical questions. What is the longest separation at which the system still gives a clean signal on the heads you will study? Can it measure short and long channels on the same probe without saturating the short ones (some systems handle this with adjustable gains or source powers)? And does the manufacturer report the noise performance in a way you can compare, for example as the noise-equivalent power, i.e., the light power that produces a signal equal to the detector’s own noise?

6. Telling Channels Apart

A detector cannot tell where the light it receives came from. It collects light from every source close enough to it, at every wavelength, together with any light in the room. To assign each part of the detected signal to the right channel and wavelength, the instrument must encode its sources, and two strategies dominate. In time multiplexing, sources (and wavelengths) are switched on one at a time in a fast, repeating cycle, so that at any given moment each detector sees only one source; often, one step of the cycle has all sources off, which measures the background light and allows it to be subtracted. In frequency multiplexing, all sources are on simultaneously, each modulated at its own frequency (typically in the kilohertz range), and the electronics separate them by frequency using lock-in or digital demodulation, much as a radio separates stations. Many instruments combine the two strategies, for example by modulating sources in frequency within groups that are switched on in turn.

Encoding also solves a problem that is easy to forget: ambient light. Room lighting, computer screens, and especially daylight contain near-infrared light, and a detector cannot distinguish it from the light of the instrument. Because the instrument’s light follows a known switching or modulation pattern, the electronics can reject light that does not follow that pattern. This rejection is not perfect, however. Strong ambient light still adds shot noise, and it can push detectors towards saturation, so shielding the optodes from direct light remains good practice, particularly outdoors (Section 7).

The outcome of the encoding scheme is the effective sampling rate of each channel, typically between about 5 and 50 Hz. In a time-multiplexed system, each additional source adds a step to the cycle, so the sampling rate per channel falls as the number of sources grows. The sampling rate must be high enough to capture the cardiac pulsation (Lesson 2.3), both because the pulsation is a useful indicator of good optode contact (Lesson 4.2) and because a signal sampled too slowly lets faster oscillations fold into lower frequencies, a distortion known as aliasing, which cannot be removed afterwards.

A 2.2 Hz heartbeat sampled at 10 Hz is captured correctly; sampled at 3 Hz, the samples trace a false 0.8 Hz oscillation

Figure 2. Aliasing. A heartbeat at 2.2 Hz, typical of an infant, is captured when each channel is sampled at 10 Hz (a). Sampled at 3 Hz (b), the same signal produces samples that trace a slow oscillation at 0.8 Hz (3 − 2.2 Hz) that does not exist in the tissue and that cannot be told apart from a real slow signal after the recording.

💭 Pause and Think

A time-multiplexed system samples each channel at 20 Hz when 8 sources are connected. You double the number of sources to 16 to cover more of the head. What happens to the sampling rate of each channel, and does it matter for your study?

Think about the frequencies you need to capture (Lesson 2.3), and about the heartbeat of an infant, which can exceed 2 Hz.

7. From Fibers to Wearables: What Really Changes

The parts of the system that touch the head are called optodes, and there are two broad ways of building them. Fiber-based systems keep the sources and detectors in a separate unit and carry light to and from the head through optical fibers or fiber bundles. Since the electronics stay away from the participant, these systems are well suited to laboratory and clinical settings, and they allow fNIRS to be recorded inside an MRI scanner. The price is weight, cable drag, and restricted movement. Fiberless systems place the sources and detectors directly on the head, with the electronics integrated into the cap or into a small unit worn on the body. They are lighter and allow the participant to move, and when they also communicate wirelessly they become truly wearable.

Wearable systems are often described simply as smaller versions of laboratory systems, but moving the instrument onto the head changes several things at once, and each has consequences for the data:

  • Motion. Removing the fibers removes cable drag, one of the main sources of motion artifacts in fiber-based systems. But participants who can move will move, and the dominant problem becomes keeping each optode in stable contact with the scalp during walking, talking, or turning the head. Cap design and optode mounting become as important as the electronics.
  • Ambient light. Outside the laboratory, and especially outdoors, ambient light can be many orders of magnitude stronger than the light returning from the head: direct sunlight delivers on the order of tens of milliwatts per square centimeter in the near-infrared, compared with the nanowatts detected at a long channel. Encoding rejects most of it, but good shielding of the optodes becomes essential.
  • Power, heat, and weight. Electronics on the head need power, produce heat, and add weight. Battery capacity limits recording time, heat must be kept away from the skin (Section 8), and the number of channels is often limited by what can be carried comfortably.
  • Timing and synchronization. A wireless device has to be synchronized with the stimulus computer and with any other devices (for example, EEG, eye tracking, or a second fNIRS system in a hyperscanning study). Wireless links add delays and occasional data loss, so the method used to mark events, and its timing accuracy, should be checked and reported.
  • Setting and design. Above all, wearable systems change what kind of study is possible: natural behavior, social interaction, classrooms, sports, and clinical settings outside the laboratory. The flip side is less experimental control, which places more weight on careful design and on recording what the participant is doing (Lesson 4.2).

🔬 Going Deeper: From the Bench to the Head

fNIRS instruments have followed a clear path from the laboratory bench to the head. The first systems measured one or a few channels, as in the early instruments for monitoring newborn infants (Cope & Delpy, 1988). Maki et al. (1995) then introduced multichannel “NIR topography,” measuring ten positions over the motor cortex simultaneously and mapping the spatial distribution of the response for the first time. Miniaturization followed: Muehlemann et al. (2008) built a miniaturized, wireless CW system, and Atsumori et al. (2009) reported a wearable, fiberless system covering the whole forehead. The current frontier is fiberless, high-density coverage of the whole scalp, which aims to bring the image quality of HD-DOT (Lesson 2.4) to wearable devices; Zhao and Cooper (2018) review the progress and the remaining obstacles.

Maki, A. et al. (1995). Medical Physics, 22(12), 1997–2005. [doi] · Muehlemann, T. et al. (2008). Optics Express, 16(14), 10323–10330. [doi] · Atsumori, H. et al. (2009). Review of Scientific Instruments, 80(4), 043704. [doi] · Zhao, H. & Cooper, R.J. (2018). Neurophotonics, 5(1), 011012. [doi]

8. Safety

fNIRS is among the safest neuroimaging methods available: near-infrared light is non-ionizing, and the powers involved are low. Low risk is not the same as no risk, however, and three aspects deserve attention in any study.

The first is skin heating. Laser and light-source safety standards define maximum permissible exposures for the skin, which depend on wavelength and exposure time, and fNIRS sources are designed to stay below them. For LEDs placed directly on the skin, however, the main risk is not the light itself but the heat conducted from the LED to the skin. Bozkurt and Onaral (2004) measured skin temperature increases of up to about 9 °C from this conducted heat, compared with less than 0.5 °C from the absorbed near-infrared light, and showed that modulating or time-multiplexing the LEDs keeps the temperature within safe limits. This is particularly relevant for long recordings, for fiberless systems with electronics on the head, and for vulnerable skin, such as that of preterm infants.

The second is the eyes. Near-infrared light is invisible, so a source can be on while it appears to be off, and the eye does not trigger its protective blink reflex. Laser sources, in particular, should never be pointed towards the eyes, fibers should not be disconnected while the sources are on, and the instrument should be operated as its documentation specifies. The third is pressure and comfort. Optodes pressed firmly against the scalp improve optical contact, but over long sessions they can cause discomfort or leave marks, especially in infants. Comfort is also a data-quality issue, since uncomfortable participants move more.

💭 Pause and Think

You are choosing between two systems for a study of toddlers playing freely in a classroom: a fiber-based system with laser sources and very sensitive detectors, and a fiberless, battery-powered system with LEDs. Which trade-offs matter most for this study?

Think about movement, weight, setup time, ambient light, synchronization with video recordings of the play, skin safety, signal quality, and what you would need to report about the light sources in your methods.


📌 Key Takeaways

  • fNIRS instruments come in four families that measure different properties of the returning light: CW (intensity; relative changes), FD and TD (travel time; absolute absorption and scattering), and speckle-based methods such as DCS (fluctuations; blood flow). CW is by far the most common.
  • A CW system records intensity time series for each channel and wavelength; hemoglobin changes are computed from them afterwards. Keep the raw intensities.
  • Only a tiny fraction of the injected light returns: about 10⁻⁶ at 3 cm in an idealized adult head model, falling by a little more than an order of magnitude per additional centimeter. Short and long channels differ by orders of magnitude, so detectors need a wide dynamic range. At standard separations, photon shot noise is far smaller than a functional response; instrument noise, ambient light, contact, and physiology set the practical limits.
  • LEDs are compact and wearable-friendly but spectrally broad; laser diodes have a well-defined wavelength and are usually fiber-coupled. Detectors trade efficiency (photodiodes) against internal gain (APDs, SiPMs, PMTs). Report source wavelengths and, for LEDs, spectra.
  • Channels are separated by time or frequency multiplexing, which also rejects most ambient light. The sampling rate per channel must capture the heartbeat to avoid aliasing.
  • Wearable systems remove cable drag but shift the challenges to optode contact during movement, ambient light, power and heat, and synchronization with other devices, while making naturalistic studies possible.
  • fNIRS is very safe, but heat from LEDs on the skin, eye exposure to invisible laser light, and optode pressure need attention, especially with infants and long sessions.

📚 Further Reading & Key References

  1. Farrell, T.J., Patterson, M.S. & Wilson, B. (1992). “A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo.” Medical Physics, 19(4), 879–888. [doi] — The diffusion model behind the light-budget estimates in Section 3.
  2. Cope, M. & Delpy, D.T. (1988). “System for long-term measurement of cerebral blood and tissue oxygenation on newborn infants by near infrared transillumination.” Medical & Biological Engineering & Computing, 26(3), 289–294. [doi] — An early CW instrument for neonatal monitoring.
  3. Maki, A., Yamashita, Y., Ito, Y., Watanabe, E., Mayanagi, Y. & Koizumi, H. (1995). “Spatial and temporal analysis of human motor activity using noninvasive NIR topography.” Medical Physics, 22(12), 1997–2005. [doi] — The first multichannel NIR topography of human brain activity.
  4. Muehlemann, T., Haensse, D. & Wolf, M. (2008). “Wireless miniaturized in-vivo near infrared imaging.” Optics Express, 16(14), 10323–10330. [doi] — An early miniaturized, wireless CW system.
  5. Atsumori, H., Kiguchi, M., Obata, A., et al. (2009). “Development of wearable optical topography system for mapping the prefrontal cortex activation.” Review of Scientific Instruments, 80(4), 043704. [doi] — An early wearable, fiberless system covering the whole forehead.
  6. Zhao, H. & Cooper, R.J. (2018). “Review of recent progress toward a fiberless, whole-scalp diffuse optical tomography system.” Neurophotonics, 5(1), 011012. [doi] — Where wearable, high-density systems are heading, and what limits them.
  7. Bozkurt, A. & Onaral, B. (2004). “Safety assessment of near infrared light emitting diodes for diffuse optical measurements.” BioMedical Engineering OnLine, 3, 9. [doi] — Skin heating from LEDs in contact with the skin, and how to limit it.
  8. Scholkmann, F., Kleiser, S., Metz, A.J., et al. (2014). “A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology.” NeuroImage, 85, 6–27. [doi] — A comprehensive review of CW instrumentation and methodology.

Society for Functional Near-Infrared Spectroscopy.
Content is for educational purposes.