Introduction to fNIRS Measuring blood flow with fNIRS
Introduction to fNIRS
0% Completed

Beyond Absorption: Measuring Blood Flow with Speckle

Every technique we have discussed so far measures how much light is absorbed, and therefore how much hemoglobin is in the tissue and how oxygenated it is. None of them measures how fast the blood is moving. This is a real gap, because blood flow is one of the three quantities that determine oxygenation, together with blood volume and oxygen consumption (Lesson 1.3), and because the neurovascular response is, at its core, a change in flow (Lesson 3.1). With absorption alone, an increase in HbO₂ can be interpreted in several ways; with flow measured independently, many of these ambiguities disappear.

In this lesson, which closes Unit 4, we look at a family of optical techniques that measure blood flow by exploiting a property of laser light that fNIRS ignores: its coherence. We start with the physical idea of speckle, then describe diffuse correlation spectroscopy (DCS), the most established of these techniques, what its blood flow index means and how it has been validated, and how it can be combined with NIRS to estimate oxygen metabolism. We finish with its limitations and with the newer variants that are trying to overcome them. As in Lesson 4.3, our aim is to help you understand what these methods measure and when they are useful, not to train you to use them, which will be the subject of a future course.

1. Speckle: What Coherent Light Reveals About Motion

If you shine a laser pointer on a wall, the spot does not look uniform: it has a grainy, sparkling texture. This pattern, called speckle, appears because laser light is coherent (its waves keep a fixed phase relationship over long distances), so light waves that reach the same point after traveling slightly different paths interfere with each other. Where they arrive in phase, they add up to a bright grain; where they arrive out of phase, they cancel into a dark one. The light that emerges from tissue after multiple scattering forms a speckle pattern of the same kind, made of the contributions of millions of photon paths.

If nothing in the tissue moves, the paths do not change and the speckle pattern is frozen. In living tissue, however, red blood cells are constantly moving, and every time a photon scatters from a moving cell, the length of its path changes by a tiny amount, comparable to the wavelength of the light. This is enough to change the interference, so the speckle pattern fluctuates, with bright grains turning dark and back again. The faster the red blood cells move, the faster the pattern fluctuates. Measuring the speed of these fluctuations is therefore a way of measuring blood flow, and, because the light has traveled through the tissue by diffusion, it measures flow at the same depths that NIRS reaches.

These techniques place specific demands on the hardware. The laser must have a long coherence length, much longer than the spread of path lengths through the tissue, which rules out the LEDs and most of the laser diodes used in CW fNIRS. The detector must see only one or a few speckle grains, since averaging over many grains washes out the fluctuations; for DCS, this means collecting light with single-mode or few-mode fibers, which capture far less light than the fiber bundles or large-area detectors of fNIRS. And the fluctuations are fast, on time scales of microseconds, so the detectors must count individual photons with fine time resolution.

2. Diffuse Correlation Spectroscopy

DCS quantifies the speed of the speckle fluctuations with the intensity autocorrelation function, \(g_2(\tau)\), which compares the detected intensity at each moment with the intensity a short delay \(\tau\) later. At very short delays, the speckle has not changed and the correlation is high; as the delay grows, the pattern decorrelates and \(g_2\) decays toward 1, its value for completely uncorrelated light. The faster the red blood cells move, the faster \(g_2\) decays. On the adult head at typical DCS separations, most of the decay happens within a fraction of a millisecond, which is why the photon arrival times must be recorded with microsecond resolution or better.

A speckle pattern, the fluctuating intensity of one speckle grain for slower and faster flow, and the corresponding autocorrelation curves

Figure 1. From speckle to blood flow. (a) Simulated speckle pattern of coherent light leaving the tissue; a DCS fiber collects light from about one grain. (b) The intensity of one grain fluctuates faster when blood flow is faster (simulated). (c) Intensity autocorrelation curves from the correlation diffusion model at 2.5 cm, for blood flow indices of 0.5 × 10⁻⁸ and 2 × 10⁻⁸ cm²/s; dotted lines mark the decay times.

To turn the decay into a measure of flow, DCS uses a diffusion model that is closely related to the one we used for NIRS. The approach has its roots in diffusing-wave spectroscopy, developed in soft-matter physics to study the motion of particles in dense, opaque suspensions (Pine et al., 1988). Boas, Campbell and Yodh (1995) showed that the temporal correlation of the light field obeys a diffusion equation of its own, the correlation diffusion equation, which can be solved in the same geometries as the photon diffusion equation of Lesson 2.5. Fitting its solution to the measured \(g_2\) curve, with the absorption and scattering coefficients of the tissue as inputs, gives a single parameter that describes how fast the scatterers move.

💡 The Blood Flow Index

The parameter recovered by DCS is the blood flow index, \(\text{BF}_i = \alpha D_B\), where \(\alpha\) is the fraction of scattering events that involve moving red blood cells and \(D_B\) is an effective diffusion coefficient describing their motion. Its units are cm²/s, not the mL/100 g/min of physiological blood flow.

Curiously, red blood cells in the microvasculature do not diffuse like Brownian particles, yet a Brownian model describes the measured curves better than a model of ordered flow. For this reason, DCS is most often used to report relative changes in blood flow, rBF = BFi / BFi,baseline, rather than absolute values.

Because the fit uses the absorption and scattering coefficients of the tissue, errors in these values propagate into the blood flow index. Many DCS studies assume them, but measuring them with an FD or TD system (Lesson 4.3) improves the accuracy, particularly when comparing baseline flow between people. Changes in absorption during a task also affect the correlation curve slightly, which is one more reason why DCS is usually combined with a NIRS measurement of the same tissue.

A fair question at this point is whether a quantity with such unusual units really tracks blood flow. Over the last two decades, relative changes in the DCS blood flow index have been compared with a wide range of established methods, including arterial spin labeling MRI, xenon-enhanced CT, transcranial Doppler ultrasound, and fluorescent microspheres, in animals and in humans, from preterm infants to adults, and have generally shown good agreement. Durduran and Yodh (2014) reviewed these validation studies together with the physical basis of the technique, and their review is the best entry point to the field.

🔬 Explore It Yourself

Change the blood flow with the preset buttons and watch the speckle flicker faster or slower and the correlation curve shift. Then change the source-detector separation: why do longer separations decorrelate faster?

🎍 MUST-READ PAPER

Durduran, T. & Yodh, A.G. (2014). “Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement.” NeuroImage, 85, 51–63.

A concise history of DCS in the brain, with an emphasis on the underlying physical ideas, the instrumentation, the validation against other blood flow measurements, and clinical and functional applications.

→ doi:10.1016/j.neuroimage.2013.06.017

3. From Flow to Metabolism: Combining DCS and NIRS

The main reason to measure flow and oxygenation together is that, with both, you can estimate how much oxygen the tissue consumes. As we saw in Lessons 1.3 and 3.1, Fick’s principle relates the cerebral metabolic rate of oxygen to blood flow and oxygen extraction: CMRO₂ = CBF × CaO₂ × OEF. If the arterial oxygen content (CaO₂) does not change during the measurement, the relative change in metabolism is simply the product of the relative changes in flow and in oxygen extraction:

$$\text{rCMRO}_2 = \text{rCBF} \times \frac{\text{OEF}}{\text{OEF}_0}$$

DCS provides rCBF. NIRS provides the oxygenation needed to estimate the OEF, usually by assuming a fixed proportion of arterial and venous blood in the tissue, so that the venous saturation, and therefore the extraction, can be inferred from the measured StO₂ or from the changes in HbO₂ and HbR. The combination was first demonstrated in the rat brain by Cheung et al. (2001), extended to imaging of blood flow, oxygenation, and metabolism during focal ischemia by Culver et al. (2003), and applied to the human brain by Durduran et al. (2004), who reported the first all-optical estimate of the relative change in CMRO₂ during sensorimotor activation. Since then, hybrid DCS and NIRS systems have been used in neonatal and adult intensive care, during surgery, and in functional studies, typically alternating between the two measurements to avoid interference between them.

💭 Pause and Think

During a motor task, a hybrid system measures a 30% increase in the DCS blood flow index, and the NIRS data indicate that the oxygen extraction fraction falls from 0.40 to 0.32. Estimate the relative change in CMRO₂. How does your answer compare with the PET results of Fox and Raichle discussed in Lesson 3.1?

Then think about which assumptions your estimate depends on: that arterial oxygen content is constant, that the ratio of arterial to venous blood is known, and that DCS and NIRS measure the same tissue volume.

4. Limitations: Photons, Depth, and the Scalp

The central limitation of DCS is the number of photons. The need to detect single speckles forces the use of small collection fibers, so a DCS detector receives only a tiny fraction of the light that a CW fNIRS detector collects at the same separation. The autocorrelation curve computed from these few photons is noisy, and obtaining a reliable blood flow index requires averaging over time, typically on the order of a second or longer for adult brain measurements. For the same reason, most adult DCS measurements use separations of about 2.5–3 cm, shorter than would be ideal for sensitivity to the brain, and DCS systems usually have far fewer channels than fNIRS systems.

Depth is a more nuanced question. Because blood flow is much higher in the brain than in the scalp and skull, DCS has a relatively higher sensitivity to the brain than NIRS at the same separation: in the simulations of Selb et al. (2014), the ratio of brain to scalp sensitivity at 3 cm was about three times higher for DCS (0.3) than for NIRS (0.1). The higher noise of photon counting, however, offset much of this advantage, and the contrast-to-noise ratio for a simulated brain activation was similar for the two methods. The scalp still contributes to the DCS signal, and it can mislead in ways that are specific to flow: pressing the probe against the head, for example, compresses the scalp and reduces its blood flow, which changes the measured signal. Short separations, pressure control, and multi-layer models are used to address this, much as in fNIRS (Lesson 3.5).

Finally, the lasers and single-photon detectors required by DCS remain more expensive than the components of a CW system, and the absolute blood flow index is difficult to compare between individuals because of its dependence on assumed optical properties and on the model. These limitations explain why DCS has so far been used mostly in specialized research and clinical settings, and they have motivated the developments described next.

5. New Directions: More Photons, More Depth

Most recent developments in speckle-based flow measurement attack the photon budget, the depth sensitivity, or both. The field is moving quickly, so the table below is not a complete list; it summarizes some of the main ideas, with an early or representative demonstration of each.

Approach Key idea What it improves
Speckle contrast (SCOS and related methods) Instead of timing the fluctuations, a camera records many speckles at once, and the blurring of the pattern during the exposure time (its reduced contrast) reveals how fast it changes (Bi et al., 2013; Valdes et al., 2014) Many speckles measured in parallel with inexpensive detectors; brain activation measured at 33 mm with over tenfold higher signal-to-noise ratio than DCS (Kim et al., 2023)
Multi-speckle DCS Arrays of single-photon detectors measure hundreds to thousands of speckles simultaneously, each with its own autocorrelation Signal-to-noise ratio grows with the number of speckles; a 32-fold gain was reported with a 1024-pixel array (Sie et al., 2020)
Interferometric DCS (iDCS) The weak light from the tissue is mixed with a strong reference beam from the same laser, which amplifies its fluctuations (Zhou et al., 2018; Robinson et al., 2020) Measurements at low light levels and longer separations, and the use of standard cameras instead of photon counters
Time-domain DCS (TD-DCS) A pulsed, coherent laser combines DCS with TD-NIRS: photons are sorted by time of flight before computing the autocorrelation (Sutin et al., 2016) Depth selectivity through late photons, and optical properties measured in the same instrument
Longer wavelengths Operating at 1064 nm instead of 765–850 nm (Carp et al., 2020) More photons per unit of energy, higher permissible exposure, and slower decorrelation, which together increase sensitivity to deep tissue

Speckle contrast methods deserve a comment, since their name is likely to appear more and more often. They grew out of laser speckle contrast imaging, a technique used since the 1980s to image blood flow on exposed surfaces, such as the cortex of animals or the retina (reviewed by Boas & Dunn, 2010). Applying the same idea to light that has diffused deep into tissue, with a point source and a detector a few centimeters away, turns it into a diffuse technique that measures the same kind of blood flow index as DCS, but with detectors that are simpler and cheaper and that capture many more speckles. Together, these developments are bringing blood flow measurements closer to the channel counts, costs, and wearability of CW fNIRS, although most of them are still in the research phase.

🔬 Going Deeper: From Colloids to the Human Brain

DCS is a good example of an idea traveling between fields. In the late 1980s, physicists studying concentrated colloids, foams, and other opaque materials developed diffusing-wave spectroscopy to measure the motion of particles in samples that scatter light many times (Pine et al., 1988). In the mid-1990s, Boas, Campbell and Yodh (1995) recognized that the temporal correlation of the light field obeys a diffusion equation analogous to the one used for photon density in tissue optics, which made it possible to measure motion deep in tissue with the same geometries and models as NIRS. The first in vivo cerebral applications followed in rodents (Cheung et al., 2001; Culver et al., 2003), and in 2004 the combination of DCS and NIRS measured blood flow, oxygenation, and oxygen metabolism in the activated human brain (Durduran et al., 2004).

Pine, D.J. et al. (1988). Physical Review Letters, 60(12), 1134–1137. [doi] · Boas, D.A. et al. (1995). Physical Review Letters, 75(9), 1855–1858. [doi] · Durduran, T. et al. (2004). Optics Letters, 29(15), 1766–1768. [doi]

6. When Is Measuring Flow Worth It?

For a typical cognitive study that asks whether a region responds to a task, CW fNIRS remains the practical choice, and adding flow measurements would bring more cost and complexity than information. Flow becomes essential when the question is about the relationship between delivery and consumption of oxygen: studies of neurovascular coupling itself, of oxygen metabolism in development, aging, or disease, and of conditions in which flow and oxygenation can change independently, such as anesthesia, changes in arterial CO₂, or impaired cerebral autoregulation. It is also valuable at the bedside, where a continuous, noninvasive measure of cerebral blood flow is otherwise hard to obtain, for example in preterm infants, in neurocritical care, or during cardiac surgery. In each of these cases, the combination of flow with absolute oxygenation from FD or TD (Lesson 4.3) gives the most complete picture that diffuse optics can currently provide.

💭 Pause and Think

A CW fNIRS study finds that HbO₂ increases less in older adults than in young adults during a cognitive task. One interpretation is that older adults activate the region less; another is that their neurovascular coupling is impaired. How could adding DCS (or another blood flow measurement) help you distinguish between these interpretations, and what would remain ambiguous?

Think about what CMRO₂ tells you that flow or oxygenation alone does not (Lesson 3.1), and about the depth and scalp issues discussed in this lesson and in Lesson 4.2.


📌 Key Takeaways

  • Absorption-based methods measure hemoglobin, not blood flow. Speckle-based methods use coherent light: moving red blood cells make the speckle pattern fluctuate, faster for faster flow.
  • DCS measures the decay of the intensity autocorrelation \(g_2(\tau)\) and fits a correlation diffusion model to obtain a blood flow index (cm²/s). Relative changes in this index agree well with established blood flow methods.
  • Combining DCS with NIRS gives relative CMRO₂ through Fick’s principle, rCMRO₂ = rCBF × OEF/OEF₀, under assumptions about arterial content and vascular compartments.
  • DCS is limited by few photons (single-speckle detection), short separations, few channels, and cost. It is relatively more sensitive to the brain than NIRS, but its higher noise offsets much of that advantage, and the scalp still contributes.
  • New approaches, including speckle contrast (SCOS), multi-speckle and interferometric DCS, TD-DCS, and longer wavelengths, increase photon counts and depth sensitivity, bringing flow measurements closer to the scale of CW fNIRS.

This lesson closes Unit 4. You now know what happens inside an fNIRS instrument, how to design and run an experiment with it, and what the more specialized members of the diffuse optics family add. In Unit 5, we turn to what happens after the recording: how raw intensities become hemoglobin changes, how artifacts and systemic physiology are handled, and how responses are estimated and compared across participants.

📚 Further Reading & Key References

  1. Pine, D.J., Weitz, D.A., Chaikin, P.M. & Herbolzheimer, E. (1988). “Diffusing wave spectroscopy.” Physical Review Letters, 60(12), 1134–1137. [doi] — The origin of correlation measurements in multiply scattering media.
  2. Boas, D.A., Campbell, L.E. & Yodh, A.G. (1995). “Scattering and imaging with diffusing temporal field correlations.” Physical Review Letters, 75(9), 1855–1858. [doi] — The correlation diffusion equation that underlies DCS.
  3. Cheung, C., Culver, J.P., Takahashi, K., Greenberg, J.H. & Yodh, A.G. (2001). “In vivo cerebrovascular measurement combining diffuse near-infrared absorption and correlation spectroscopies.” Physics in Medicine and Biology, 46(8), 2053–2065. [doi] — Combined DCS and NIRS in the rat brain during hypercapnia.
  4. Culver, J.P., Durduran, T., Furuya, D., Cheung, C., Greenberg, J.H. & Yodh, A.G. (2003). “Diffuse optical tomography of cerebral blood flow, oxygenation, and metabolism in rat during focal ischemia.” Journal of Cerebral Blood Flow & Metabolism, 23(8), 911–924. [doi] — Imaging of flow, oxygenation, and CMRO₂ in a stroke model.
  5. Durduran, T., Yu, G., Burnett, M.G., et al. (2004). “Diffuse optical measurement of blood flow, blood oxygenation, and metabolism in a human brain during sensorimotor cortex activation.” Optics Letters, 29(15), 1766–1768. [doi] — The first all-optical estimate of relative CMRO₂ in the human brain.
  6. Durduran, T. & Yodh, A.G. (2014). “Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement.” NeuroImage, 85, 51–63. [doi] — History, physics, validation, and applications of DCS in the brain.
  7. Selb, J., Boas, D.A., Chan, S.-T., Evans, K.C., Buckley, E.M. & Carp, S.A. (2014). “Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia.” Neurophotonics, 1(1), 015005. [doi] — Brain versus scalp sensitivity and contrast-to-noise of DCS and NIRS.
  8. Boas, D.A. & Dunn, A.K. (2010). “Laser speckle contrast imaging in biomedical optics.” Journal of Biomedical Optics, 15(1), 011109. [doi] — A review of speckle contrast imaging, the precursor of diffuse speckle contrast methods.
  9. Bi, R., Dong, J. & Lee, K. (2013). “Deep tissue flowmetry based on diffuse speckle contrast analysis.” Optics Letters, 38(9), 1401–1403. [doi] — Deep-tissue flow from the spatial statistics of speckle recorded with a camera.
  10. Valdes, C.P., Varma, H.M., Kristoffersen, A.K., Dragojevic, T., Culver, J.P. & Durduran, T. (2014). “Speckle contrast optical spectroscopy, a non-invasive, diffuse optical method for measuring microvascular blood flow in tissue.” Biomedical Optics Express, 5(8), 2769–2784. [doi] — Introduced SCOS, combining speckle contrast with the photon diffusion model.
  11. Sutin, J., Zimmerman, B., Tyulmankov, D., et al. (2016). “Time-domain diffuse correlation spectroscopy.” Optica, 3(9), 1006–1013. [doi] — Depth-resolved blood flow and optical properties from time-gated DCS.
  12. Zhou, W., Kholiqov, O., Chong, S.P. & Srinivasan, V.J. (2018). “Highly parallel, interferometric diffusing wave spectroscopy for monitoring cerebral blood flow dynamics.” Optica, 5(5), 518–527. [doi] — Interferometric detection with a standard camera, measuring pulsatile blood flow in the adult brain.
  13. Robinson, M., Boas, D.A., Sakadžić, S., Franceschini, M.A. & Carp, S.A. (2020). “Interferometric diffuse correlation spectroscopy improves measurements at long source-detector separation and low photon count rate.” Journal of Biomedical Optics, 25(9), 097004. [doi] — Characterizes the gains of interferometric (heterodyne) DCS.
  14. Carp, S.A., Tamborini, D., Mazumder, D., et al. (2020). “Diffuse correlation spectroscopy measurements of blood flow using 1064 nm light.” Journal of Biomedical Optics, 25(9), 097003. [doi] — The advantages of longer wavelengths for deep-tissue DCS.
  15. Sie, E.J., Chen, H., Saung, E.-F., et al. (2020). “High-sensitivity multispeckle diffuse correlation spectroscopy.” Neurophotonics, 7(3), 035010. [doi] — Parallel detection of many speckles with a single-photon detector array.
  16. Kim, B., Zilpelwar, S., Sie, E.J., et al. (2023). “Measuring human cerebral blood flow and brain function with fiber-based speckle contrast optical spectroscopy system.” Communications Biology, 6, 844. [doi] — Speckle contrast measurements of human brain activation at long separations.